OPTIMA MR360 & BRIVO MR355
FULL SERVICE TRAINING REV 1.5
ATC MR INSTRUCTOR XIAO XIAO
GE ASIA TRAINING CENTER
Content
Module 1 - Course Introduction & Safety..............................................................1
Module 2 - System overview………………………………………………………………………….14
Module 3 - New User Interface and Software Operation……………………...65
Module 4 – Architecture & Hardware…………………………………………………………..87
Module 5 – Calibration & Diagnostics……………………………………………………….169
Lab Guide & Component Identification…………………………………………………...193
1
2
3
4
Excite HD training is a pre-requisite for this course. If you haven’t been trained
on Excite HD, it is mandatory to get the HD training before you will be able to
be successful in the SV course.
5
6
7
There are three types of safety hazards to be aware of with MR. Always
observe high voltage dangers and perform Lock Out / Tag Out (LOTO) safety
steps before working on any piece of MR equipment. Do not substitute a
shortcut process for LOTO. A shortcut that produced one set of desired results
on an older piece of equipment may not produce the same set of desired
results on the new piece of equipment.
Complete the required MR512 MR Magnet and Cryogen Safety course before
attempting to service a magnet or work with cryogens.
Note to instructors: You can embellish as you wish since you both use
different slides for safety in your classes.
8
Be sure to be cautious when installing or replacing anything in the magnet
room, especially in the 3.0T environment. Test items first for ferrous properties.
Never put your body between the magnet and anything that may be a ferrous
material.
The 3T magnetic field requires extreme caution. Be alert for ferrous materials
when performing replacement procedures in the scan room. Note that on 3T
systems, even non-ferrous tools are slightly ferrous! Follow all service
documentation procedures.
The System Pre-Installation manual shows magnetic field strengths at
different distances from the magnet
9
The portable GasWatch2 Oxygen sensor as pictured above - part number is
5111049
10
Due to the magnet top off you may have to sting the magnet and handle
dewars during this fill. The gas cylinders must be strapped down and when not
in use have their cap on, also be very careful of the dewars. In most cases, Air
Products or a cryogen contractor will take care of this task, but you should be
aware of this in case you need to perform it .
Frostbite:
•Distance and Time- Frostbite Protection
11
LOTO Service Documentation is Accessible via the Service Methods Website by
clicking on “Safety” and then the correct procedure name under “Lock Out Tag
Out”. Note that for SV, all LOTO procedures involving LOTO of the SYSTEMS
CABINET require shut down of the Simple OC first . This is because powering
down the Simple OC ensures correct soft power down of the Image Compute
Node. The ICN MUST be shut down via a soft power down in order to avoid
operating system corruption!
12
13
14
15
16
17
Compared to the HD system, SV changes lots of components.
• Air cooling system is replaced by water cooling system for RF amplifier and Gradient
amplifier.
• System cabinet, HFD cabinet and part of penetration panel are combined in one
compact cabinet.
• Components in the compact cabinet are new designed except for GP3, Term Server and
HUB
• Equipment in Operator room is simplified
• Fixed table with the express coil in the cradle will relieve the operator ’s labor for
changing coils
• Three parts in Magnet side kit are new designed for SV
18
19
20
If you refer to the MR Service Docs for
5364191-2EN - Optima MR360 / Brivo MR355 1.5T Service Methods>Replacement>Magnet
Enclosure>Magnet Enclosure Table of Contents
you will see the same diagram with clickable links for each component.
These links will take you to the procedures for removing and reinstalling the covers.
21
Refer to the service manual for detail procedure of replacement:
5364191-2EN - Optima MR360 / Brivo MR355 1.5T Service Methods>Replacement>Magnet
Enclosure>Magnet Enclosure Control Panels
SV, HDe, HDx Forward Production Control Panels:
• Same function as control panel in HD except for the appearance
• Now have words written next to the buttons
• Languages specific control panels are shipped with the Keyboard collectors in above-
mentioned systems
• May be removed independently
• Are secured with poppers and attached to hidden cables
• Be sure to follow all procedures in MR Service Documentation when removing any cover
• Note - Top Laser Light is now housed inside the top display panel. Use caution when
popping the panel off to adjust this laser so as not to trip the Emergency Stop
22
Important facts about the LPCA of SV:
• There is no head T/R switch. T/R switch is moved to the Quick disconnect of the Split head
coil and Quad knee coil.
• Port A is a coil interface for coils with and without preamps
• Each receive line has single input for coils with or without preamps. Preamp or non-
preamp path is controlled by shorting pins on Port A in the coil connector
• Head Tx RF is routed through the interface module, this is done to “tap-off” the head T/R
bias
• Head T/R bias is used for preamp protect switching in the interface module
• Multi Coil bias is added to the RF coax lines in the port connectors
• Bias is added to Port A in the Multicoil Interface Box
• RF input to preamp boards is routed though coax cable
23
Rear Pedestal is newly designed.
1. The 16ch Switch is not used in SV system, which is functionally replaced by Mega Switch
located on the Magnet Side kit.
2. LED Power Box is introduced, which is the power supply for SRI, PAC, Patient alignment
laser and bore light.
3. Dummy load is moved to rear pedestal from the LPCA.
4. Hybird, TNS antenna, motor, encoder assembly, encoder box and communication box
have no change. They are same as the parts in HD system.
24
• Magnet side kit contains Voltage Regulator Box, Mega Switch, RRX, cables and base plate.
It is mounted on the left side of the magnet through insulating board, and a ground wire
is connected between base plate and system cabinet.
• Voltage Regulator Box and Mega SW are both newly designed components. Regulator
Box regulates the DC power from DCPS for Mega SW & RRX. Mega SW combines the
function of 8ch switch, MUX, UTNS and receiver. RRX used in SV system is 8 channels.
• Mega Switch select different input RF signal, and amplifies them according to R1 value,
and transform the center frequency of RF signal to 16MHz, then send them to RRX. RRX
converts the analog RF signal to digital signal and transfer them to VRF in ICN by optical
fiber.
25
• Fixed table has no Dock and Undock function like detachable table. Up and down of fixed
table are implemented by actuator, which is driven by electrical motor. The table can go
up and down very smoothly with very low noise compared to the detachable table. And
there is no other mechanism to drive fixed table up and down.
• The control box for fixed table is newly designed.
• The longitudinal movement of the cradle is driven by longitudinal motor located in the
rear pedestal. It is same as HD.
• The cover of the cradle is reinforced to protect the embedded Express PA coil from
damage. The cradle with the Express PA coil moves the patient into and out of the
magnet bore.
26
Express coil consists of three parts:
1) HNA, functionally similar to NV coil;
2) PA, functionally similar to CTL coil;
3) AA, functionally similar to 8Ch body array or 4Ch torso coil.
HNA and AA are removable coils that connected to Port A in LPCA. PA coil is embedded in
the cradle of the fixed table, which can be moved in and out of the magnet bore with the
cradle. The interface board of PA coil located on the bottom of the cradle. Two cables
always connect the PA interface
,the suitable mode according to the landmark position. This function is only
available for Express coil, and not all the scan modes of the Express coil are compatible
with this function. This table shows all the available modes.
• This two pictures show the procedure to use this function. In step 4, after selection of the
automatic coil selection, the background color of one configuration in the right area will
turn yellow, which means this mode has been selected automatically.
167
Please refer to the service method 5364191-2EN 'Replacement / Fix Table / Express Coil:
Posterior Array Coil Replacement without removing Fixed Table' for the detailed procedure.
168
169
170
171
It is important during new installations and upgrades, to verify that your IP addresses are
always up-to-date (both Host IP and MGD IP). If any changes have been made to an IP
address, you must reconfigure the the ICN.
172
173
There are some steps during the installation of the MR system that must be done but for
which the system has no “machine” data. For these items the system will rely on the user to
check off the items as they are complete and click [Submit].
For all calibrations that have machine data, the tool can be launched the same way it was
launched in previous versions of MR software – by clicking [Click here to start this tool]. The
tool will launch in a separate window. After completion of the tool, the user must click
[Check Status].
After clicking [Submit] or [Check Status], the ICW will check the data files. If it finds a passed
calibration or a completed checklist, then it will allow the user to advance to the next
calibrations in the flow. If it detects a failure or an incomplete task, it will turn the name of
the tool RED in the left window of the ICW and no new calibrations will become available
until that step has been completed successfully.
174
175
How to read this matrix: The same tasks calibrations are listed along the top and along the
side. The ideal order of task / calibration completion is from the top to the bottom. Anything
with no red squares to the right has no pre-dependencies, therefore it can be run at any time.
If there are red squares to the right of any of the tasks/calibrations, then the
tasks/calibrations indicated in the vertical column(s) associated with each red square must
be completed successfully before that task/calibration becomes available.
This matrix is the basis for the logic of the ICW
176
Every time the user clicks on the “Calibration” tab of the CSD, as long as there is a service
key plugged in, the ICW will be the only link available. Once all steps in Install Mode have
been successfully completed, Maintenance Mode becomes available.
In Maintenance Mode, any time you click on a calibration it will alert you to the possibility of
pre- and post- dependencies. Choosing [Yes] acknowledges understanding of these
dependencies and pre-dependencies will be de-bolded. Clicking [No] de-selects calibration.
Clicking [More] shows the Calibration Dependency Matrix.
177
This matrix is the basis for the logic of the ICW in Maintenance Mode. If you click [More] on
the pop-up from the previous slide, this is the matrix that you will see. It is good to become
familiar with how to read this matrix so that when you perform re-calibrations at sites you
will understand what it means.
Maintenance Mode Dependency Matrix:
How to read this matrix: The same tasks are listed vertically and horizontally. In the column
on the left find the calibration that you wish to run. To the right of that calibration name,
follow the line horizontally until you reach the yellow square. Any task vertically directly
above the yellow square is a pre-dependency, and any task vertically directly below the
yellow square is a post-dependency.
Post-dependencies are more important than pre-dependencies. For troubleshooting
purposes, if user is about to perform a certain calibration (or task), it would be beneficial to
see what pre-dependencies are related to that calibration and consider performing them. If
pre-dependencies are not run, the system will not be negatively affected. Post-
dependencies, however, are very important. For example, if Grafidy is run, LVShim, Gradcal,
EPT and Probe values may have changed. Each one of these calibrations should be run
after the completion of Grafidy because they are Post-Dependencies of that cal.
178
179
Before using the software to do the LVShim, LVshim localizer scan must be done to assure
the phantom is in the center.
180
Insert appropriate DQA-III Phantom into head coil. Verify phantom is inserted into head coil.
Also verify phantom is not rotated within the head coil.
The correct phantom type, position and landmark are all critical to successful operation of
the DQA-II tool.
If there is a failure during the DQA-II tool, the “Actions Required” window will give details as
to the likely reason for the error. If specific recommendations are not available in the
“Actions Required” window, the FE will be pointed to Service Documentation.
181
This new style 6-channel Grafidy fixture couldn’t be easier to use! It is simply a “plug and go”
tool, meaning that when it is taken out of the box all that is required is that the user attaches
the adaptor cable to the fixture and plugs it into the LPCA A-Port connector, unfolds the
wings of the fixture, landmarks and begins the Grafidy calibration. With the new fixture, it is
not necessary to connect cables from the fixture through the pen panel, nor is it necessary
to attach power supplies to the Systems cabinet or attenuators to the RF Amplifier. In
addition, once the landmark is established, no further user interaction with the Grafidy
fixture is required. This new 6-channel Grafidy fixture will be available in both 1.5T and 3.0T
versions and will be available as a service tool for all SV systems. Previous versions of
Grafidy fixtures will not work with SV software, and at this point the new 6-channel Grafidy
fixture will only function on systems running after 14x software.
182
Along with the new 6-channel Grafidy fixture defined in the previous slide, improvements
have been made to the Grafidy calibration’s data acquisition and software. The tool first
verifies proper fixture placement and signal from all channels. It then acquires data
simultaneously from all channels, thus reducing calibration time. With the new fixture and
new Grafidy software, Grafidy calibrations take about 45 minutes on BRM and CRM systems
and about 1.5-2 hours on TRM systems. In addition, the user interface has been modified.
In Manual Mode, the UI has changed from “coil placement” centered to “applied gradient”
because now one gradient can be pulsed to get data from all six coils
instead of just two as in the previous Grafidy releases. For 1.5T, as compared to previous
Grafidy, with the new software, using data from all six coils together allows for more
accurate B0 component. Also, after changing calibration factors, the UI does not allow the
user to exit without collecting data, and the number of significant digits displayed has been
reduced to make the values more readable. Grafidy can now compensate very long time
constants more accurately. The new Grafidy software allows for much better compensation
of the initial part of the eddy current curve in long mode because of three things: 1) there is
an additional time constant used 2) all data is now used (as opposed to data starting at
2.5msec before) and 3) software allows for negative amplitude.
183
184
185
186
The report file is stored in /export/home/signa/bin.
187
188
189
The Multicoil Receive Chain Tool provides a means of diagnosing problems in Multicoil
Receive (MCR) chain hardware, without use of coils. The tool sends signals down the
individual paths to isolate specific channels in the receive chain. The
,tool’s Graphical User
Interface (GUI) guides the user through the various steps involved in running the tool and
troubleshooting the problem. The GUI has built-in instructions and detailed setup instructions,
troubleshooting flowcharts and documents that will help facilitate rapid diagnosis of the
problem. There are two levels of tests that can be run using the MCR III tool: Default and
Individual. Once you've decided to use MCR III tool, determine which level of test your
situation requires. The default and port-specific levels of tests offer increased user-friendly
automation that is been omitted from the individual tests by design. Therefore, it is
imperative to identify which level of tests you need to perform and then follow the
instructions for that specific level.
1) Port A Default Tests
This test must be passed during the MR installation.
In most scenarios, this is the first test that should be run when using the MCR tool.
If this test fails, individual tests can be run to further isolate the failing component.
2) Port A Individual Tests
These tests are designed to run after the default test indicates a failure.
3) Express Coil Test For Fixed Table
This test must be passed during the MR installation.
190
The RF and DC cables for Express PA Coil - Cradle mode are shipped with the system.
191
Optima MR360 / Brivo MR355
GEHC-TECH-MD33392427
MR5047
Service Training
Lab Guide
Rev Number: J
MyWorkshop Book Number: 51232
GE Healthcare MR5047 SV Lab Guide
Page: 1 of 24
Table of Contents
TABLE OF CONTENTS.......................................................................................................................... 1
1. SYSTEM GEOGRAPHY LAB .............................................................................................................. 2
2. OPERATION LAB .............................................................................................................................. 4
3. CAN LINK LAB ................................................................................................................................. 5
4. DVMR LINK LAB .............................................................................................................................. 7
5. ICN DIAGNOSTIC LAB ................................................................................................................. 11
6. SIMPLE OC COMPUTER AND DUST FILTER REPLACEMENT LAB .......................................... 13
7. COOLANT PUMPOUT LAB .......................................................................................................... 14
8. XFD REPLACEMENT LAB ............................................................................................................. 15
9. CABINET MONITOR LAB ............................................................................................................. 17
10. FIXED-TABLE ACTUATOR LAB ................................................................................................. 19
11. FIXED-TABLE INSTALLATION & ADJUSTMENT LAB ............................................................. 20
12. EXPRESS COIL LAB .................................................................................................................... 22
13. MCR III TOOL LAB ...................................................................................................................... 23
14. MODIFIED CALIBRATION TOOL LAB ...................................................................................... 24
GE Healthcare MR5047 SV Lab Guide
Page: 2 of 24
1. System Geography Lab
Description: The purpose of this lab is to investigate and identify the major
components of the MR360 / MR355 System.
Time Required: 1 Hour
Learning
Objective(s): Identify the major subsystems and functions of the SV system
Support Materials
Required:
Service Methods procedures – 5364191, on disk, or as loaded onto
system computer
Tools Required: Handout as provided
Safety
Considerations:
Follow all LOTO procedures. They are posted in the lab as well as
incorporated as subset of documented procedure.
Steps:
1. System Power-Down
Follow software and power shutdown procedure in Service
Methods 5364191
a. Shut down applications and turn system power off.
2. System Components Geography
Using the handout provided:
a. Match the subsystems and component names to the boxes
indicating their location. Open cabinet doors for interior
identification
b. Remove covers as needed, for interior identification
3. For Simple Operator Console
a. Remove cover and expose right side components to view
location.
Do not disconnect any cabling
4. For Magnet
a. Remove cover and expose left side components to view
locations
Do not disconnect any cabling
5. For Penetration Panel
a. Remove access panel from scan room side to view locations.
Do not disconnect any cabling
6. For Rear Pedestal
a. View component locations
Do not disconnect any cabling
GE Healthcare MR5047 SV Lab Guide
Page: 3 of 24
7. For Fixed Table
a. View component locations
Do not disconnect any cabling
8. For Water Chiller
a. View component locations
Do not disconnect any cabling
9. System Power-On
a. Follow power and software restoration procedure in Service
Methods 5364191
If subsequent labs are to be performed – you may be instructed to
leave covers off. Please ensure covers are properly and safely stored
to avoid trip hazards.
Review:
Questions for instructor:
______________________________________________________
______________________________________________________
______________________________________________________
GE Healthcare MR5047 SV Lab Guide
Page: 4 of 24
2. Operation Lab
Description: The purpose of this lab is to understand basic operation procedures
of the MR355/360 scanner.
Time Required: 2 Hour
Learning
Objective(s):
Operate the system, as outlined in the service documentation.
Verify system settings and performance.
Support Materials
Required:
MR355/360 system capable of being powered up to normal
scanning condition.
Most recent revision of MR360/355 Operator Manual, on disk,
online, or as loaded onto system computer.
(http://3.28.123.26:8000/mr/OPEMAN/HDsv/M3/index.htm)
Tools Required: Laptop
Safety
Considerations:
Follow all LOTO procedures as required. They are posted in the lab
as well as incorporated as subset of documented procedure.
Steps:
1. Phantom Preparation
a. Using documented procedure in MR355/360 Operator
Manual: Patient Preparation. Set up the phantom for
scanning.
2. Prescan
a. Using documented procedure in MR355/360 Operator
Manual: Prescan. Start Auto Prescan and Manual Prescan.
3. Scan
a. Using documented procedure in MR355/360 Operator
Manual: Scan. Start to scan.
4. View
a. Using documented procedure in MR355/360 Operator
Manual: Viewer. View the images
5. Save / Restore images
a. Using documented procedure in MR355/360 Operator Manual:
Img Mgmt Data Apps. Save and restore the images
6. Add Remote Host
a. Using documented procedure in MR355/360 Operator Manual:
Img Mgmt Archive Network. Add remote host
7. Configure / Add Film Printer
a. Using documented procedure in MR355/360 Operator Manual:
Film. Add film printer
Review:
Questions for instructor:
______________________________________________________
GE Healthcare MR5047 SV Lab Guide
Page: 5 of 24
3. CAN link Lab
Description: This lab will familiarize the student with the CAN Link Diagnostic
Time Required: 1 Hour
Learning
Objective(s):
Navigate the service documentation
Understand how the CAN Link works in SV system
Perform the CAN Link diagnostic and issue isolation procedure
Support Materials
Required:
Service Methods procedures
,– 5364191 on disk, or as loaded onto
system computer. Refer to CAN Link Diagnostic procedure.
(http://3.28.123.26:8000/mr/cdrom/5364191-
2EN/root/data/Signa_EXCITE/content/1084467.htm)
Tools Required: Standard hand tools
Safety
Considerations:
Follow all LOTO procedures. They are posted in the lab as well as
incorporated in the documented procedures
Steps:
1. CAN Link Diagnostic
Use documented procedure in Direction 5364191:
Troubleshooting / Diagnostics / System Function / Patient
Handling / CAN Link Diagnostic
OR
Troubleshooting / Diagnostics / Hardware Location / Misc /
CAN Link Diagnostic
a. Perform CAN Link Diagnostic
2. Refer to the following CAN link diagram to perform the test:
a. Remove the terminator from cabinet monitor
b. Perform step 1 and record the test result
_______________________________________________________
c. Remove the connector J19 on the front panel of SRFD3 and
install the terminator on the front panel
d. Perform step 1 and record the test result
_______________________________________________________
d. Remove the connector J2 on the front panel of Driver Module
Lite and install the terminator on the front panel
e. Perform step 1 and record the test result
_______________________________________________________
f. Restore the connectors J2 on Driver Module Lite and terminator
on J11 of Cabinet Monitor
i. Remove connectors J18 and J19 from SRFD3, and connect them
together
GE Healthcare MR5047 SV Lab Guide
Page: 6 of 24
j. Perform step 1 and record the test result
_______________________________________________________
k. Remove connectors J1 and J2 from PHPS Lite2, and connect
them together
l. Perform step 1 and record the test result
_______________________________________________________
m. Restore all the cables
Questions:
1. What is the function of force reset button?
________________________________________________________
2. What conclusion can be made according to step 2?
_______________________________________________________
_______________________________________________________
______________________________________________________
Review:
Questions for instructor:
_______________________________________________________
GE Healthcare MR5047 SV Lab Guide
Page: 7 of 24
4. DVMR Link Lab
Description: This lab will familiarize the student with the DTX board level
diagnostics and DTX to RRX data path
Time Required: 2 Hour
Learning
Objective(s):
Navigate the service documentation and perform board and data
path diagnostics.
Identify error messages associated with certain failure modes
Understand the real time diagnostic function of DVMR link
Support Materials
Required:
Service Methods procedures – 5364191 on disk, or as loaded onto
system computer.
Tools Required: Standard hand tools
Safety
Considerations:
Follow all LOTO procedures. They are posted in the lab as well as
incorporated in the documented procedures
Steps: 1. DTX Board Level Diagnostics
Use documented procedure in Direction 5364191:
Troubleshooting/Diagnostics/System Function/RF/Transmitter
Diagnostics
a. Perform DTX board Level Diagnostics
b. Record results: ____________________________________
c. Place the RF Enable switch in the front panel of exciter in the
Disable (Down) position
d. Perform DTX board Level Diagnostics
e. Record results: ____________________________________
f. Use command mgd_term to open SCP and AGP terminal
g. Do the TPS Reset and check what will happen
h. Restore the RF Enable switch to Up position
2. RF Calib
Use documented procedure in Direction 5364191:
Troubleshooting/Diagnostics/System Function/RF/Transmitter
Diagnostics
a. Perform RF Calib
b. Record results: ____________________________________
c. Disconnect RUN M2502 from J107 of Mega Switch
d. Perform RF Calib
e. Record results: ____________________________________
f. Use command mgd_term to open SCP and AGP terminal
g. Do the TPS Reset and check what will happen
h. Restore the cable
i. Disconnect RUN M1514 from J105 of Mega Switch
j. Perform RF Calib
GE Healthcare MR5047 SV Lab Guide
Page: 8 of 24
k. Record results: ____________________________________
l. Use command mgd_term to open SCP and AGP terminal
m. Do the TPS Reset and check what will happen
n. Restore the cable
o. Disconnect RUN M1515 from J103 of Mega Switch
p. Perform RF Calib
q. Record results: ____________________________________
r. Use command mgd_term to open SCP and AGP terminal
s. Do the TPS Reset and check what will happen
t. Restore the cable
3. DTX -> RRX DP
Use documented procedure in Direction 5364191:
Troubleshooting/Diagnostics/System Function/RF/Transmitter
Diagnostics
a. Perform DTX -> RRX DP Diagnostics
b. Record results: ____________________________________
c. Disconnect RUN M1513 from J106 of Mega Switch
d. Perform DTX -> RRX DP Diagnostics
e. Record results: ____________________________________
f. Restore the cable
4. Observe Error Message
Refer to the following diagram
a. Record the status of LEDs on the IRF3 board and RRX
b. Disconnect IRF3 J8
c. Note the status of LEDs on the IRF3 board and RRX
d. Record error code: ____________________________________
e. Restore IRF3 J8
f. Disconnect IRF3 J11
g. Note the status of LEDs on the IRF3 board and RRX
J21
To VRF
GE Healthcare MR5047 SV Lab Guide
Page: 9 of 24
h. Record error code: ____________________________________
i. Restore IRF3 J11
j. Disconnect IRF3 J13
k. Note the status of LEDs on the IRF3 board and RRX
l. Record error code: ____________________________________
m. Restore IRF3 J13
n. Disconnect RRX J4
o. Note the status of LEDs on the IRF3 board and RRX
p. Record error code: ____________________________________
q. Restore RRX J4
r. Disconnect Exciter J21
s. Note the status of LEDs on the IRF3 board and RRX
t. Record error code: ____________________________________
u. Restore Exciter J21
5. IRF3 Wide Coverage Diagnostics
Use documented procedure in Direction 5364191:
Troubleshooting/Diagnostics/System Function/RF/ IRF3 Wide
Coverage Diagnostics
a. Perform IRF3 Wide Coverage Diagnostics
b. Replace the Fiber-optic Transceiver on IRF3 J10 with a broken
one, and perform IRF3 Wide Coverage Diagnostics
c. Record results: ____________________________________
d. Restore the good Fiber-optic Transceiver
e. Replace the Fiber-optic Transceiver on IRF3 J11 with a broken
one, and perform IRF3 Wide Coverage Diagnostics
f. Record results: ____________________________________
g. Restore the good Fiber-optic Transceiver
h. Replace the Fiber-optic Transceiver on IRF3 J13 with a broken
one, and perform IRF3 Wide Coverage Diagnostics
i. Record results: ____________________________________
j. Restore the good Fiber-optic Transceiver
k. Replace the Fiber-optic Transceiver on Exciter J19 with a
broken one, and perform IRF3 Wide Coverage Diagnostics
l. Record results: ____________________________________
m. Restore the good Fiber-optic Transceiver
n. Replace the Fiber-optic Transceiver on RRX J4 with a broken
one, and perform IRF3 Wide Coverage Diagnostics
o. Record results: ____________________________________
p. Restore the good Fiber-optic Transceiver
q. Replace the Fiber-optic Transceiver on RRX J5 with a broken
one, and perform IRF3 Wide Coverage Diagnostics
r. Record results: ____________________________________
s. Restore the good Fiber-optic Transceiver
t. Replace the Fiber-optic Transceiver on VRF with a broken one,
GE Healthcare MR5047 SV Lab Guide
Page: 10 of 24
and perform IRF3 Wide Coverage Diagnostics
u. Record results: ____________________________________
v. Restore the good Fiber-optic Transceiver
Questions: 1. What diagnostics are performed during IRF3 Wide Coverage
Diagnostic?
,______________________________________________________
______________________________________________________
______________________________________________________
GE Healthcare MR5047 SV Lab Guide
Page: 11 of 24
5. ICN Diagnostic Lab
Description: This lab will familiarize the student with the function of three
ethernet cables connected to ICN.
Time Required: 1.5 Hour
Learning
Objective(s): Understand the function of NET MGT, Port2 and Port 3 of ICN
Support Materials
Required:
Service Methods procedures – 5364191 on disk, or as loaded onto
system computer. Refer to ICN Reconfiguration procedure.
Tools Required: Standard hand tools
Safety
Considerations:
Follow all LOTO procedures as required. They are posted in the lab
as well as incorporated as subset of documented procedure.
Steps:
1. Understanding Ethernet cables function
a. Open CShell window and input more /etc/hosts
b. Record the IP address of icn1 vre and icn1_bmc
c. Start a new Exam to scan
d. Disconnect the cable connected to NET MGT port of the ICN
e. Open a C-shell window and ping icn1 and icn1-bmc.
f. Start to scan
g. Disconnect the cable connected to Port2 of the ICN
h. Ping icn1 and icn1-bmc
GE Healthcare MR5047 SV Lab Guide
Page: 12 of 24
i. Start to scan
j. Restore the disconnected cable to Port2 of the ICN
k. Disconnect the cable connected to Port3 of the ICN
l. Ping icn1 and icn1-bmc
m. Start to scan
n. Disconnect the cable connected to Port2 of the ICN again
o. Ping icn1 and icn1-bmc
p. Start to scan. Observe what will happen and record error
message
______________________________________________________
q. Restore all the disconnected cables
Questions:
1. What are IP addresses of the NET MGT Port, Port2 and Port3?
_______________________________________________________
2. What data are transferred thru NET MGT Port, Port2 and Port3?
_______________________________________________________
Review:
Questions for instructor:
______________________________________________________
______________________________________________________
______________________________________________________
GE Healthcare MR5047 SV Lab Guide
Page: 13 of 24
6. Simple OC Computer and Dust Filter Replacement Lab
Description: This lab will familiarize the student with the procedure for replacing
the OC host computer
Time Required: 1.5 Hour
Learning
Objective(s):
Navigate the service documentation
Remove and replace Simple OC computer
Support Materials
Required:
Service Methods procedures – 5364191 on disk, or as loaded onto
system computer. Refer to Simple OC and Dust Filter replacement
procedure.
(http://3.28.123.26:8000/mr/cdrom/5364191-
2EN/root/data/Signa_EXCITE/content/1106757.htm)
Tools Required: Standard hand tools
Safety
Considerations:
Follow all LOTO procedures. They are posted in the lab as well as
incorporated in the documented procedures
Steps:
1. Simple OC Computer Replacement
Use documented procedure in Direction 5364191:
Replacement /Operator Workspace Replacement / Computer
Replacement
a. Perform steps 1 – 12
b. DO NOT remove the disk drive!
2. Dust Filter Replacement
Use documented procedure in Direction 5364191:
Replacement /Operator Workspace Replacement / Dust Filter
Replacement
a. Perform Dust Filter replacement
Review:
Questions for instructor:
______________________________________________________
______________________________________________________
______________________________________________________
GE Healthcare MR5047 SV Lab Guide
Page: 14 of 24
7. Coolant Pumpout Lab
Description: This lab will familiarize the student with the procedure to pump out
coolant from SRFD3 and XFD in SV system.
Time Required: 0.5 hour
Learning
Objective(s):
Navigate the service documentation
Understand the importance of coolant pumpout before
components replacement
Practice the correct procedure of coolant pumpout
Support Materials
Required:
Service Methods procedures – 5364191 on disk, or as loaded onto
system computer. Refer to Coolant Pumpout procedures.
(http://3.28.123.26:8000/mr/cdrom/5364191-
2EN/root/data/Signa_EXCITE/content/1093487.htm)
Tools Required: Coolant Tank, Water tray, Hand Pump and Standard hand tools
Safety
Considerations:
Follow all LOTO procedures as required. They are posted in the lab
as well as incorporated as subset of documented procedure.
After draining, the drain valves must be closed. Check that drain
valves are closed.
Avoid coolant splashing into the electronics.
Steps: 1. Coolant pumpout
Use documented procedure in Direction 5364191:
Replacement / System Cabinet Replacement / XFA and XFD-PS
Replacement / XFD-PS / 3 Procedure / 3. Drain the coolant
a. Perform the draining operation
Questions: 1. What is the use of water tray?
__________________________________________________________
2. Why the coolant in the water-cooling components must be drained
when replacement?
__________________________________________________________
3. What action should be executed before restoring the drain hose to
drip pan?
__________________________________________________________
Review: Questions for instructor:
______________________________________________________
______________________________________________________
______________________________________________________
GE Healthcare MR5047 SV Lab Guide
Page: 15 of 24
8. XFD Replacement Lab
Description: This lab will familiarize the student with the procedure to replace the
XFA and XFD-PS
Time Required: 2 hours
Learning
Objective(s):
Navigate the service documentation
The hoist tool
Practice the correct procedure of XFD replacement
Know how to pull out the XFD in case XFD can’t be pulled out
according to the normal procedure
Support Materials
Required:
Service Methods procedures – 5364191 on disk, or as loaded onto
system computer. Refer to XFD replacement procedures.
(http://3.28.123.26:8000/mr/cdrom/5364191-
2EN/root/data/Signa_EXCITE/content/1084772.htm)
Tools Required: Hoist tool, Hoist bracket for XFD, Standard hand tools and DVM.
Safety
Considerations:
Follow all LOTO procedures as required. They are posted in the lab
as well as incorporated as subset of documented procedure.
Discharging the super capacitor by waiting for at least 15 minutes
after system power off for XFD replacement, or at least 30 minutes
for removing power panel assy.
The weight of XFD-PS is 87kg. Take care when replacement.
The left top of XFD-PS will be over 40oC in ten minutes after
discharging. Take care when replacement.
Steps: 1. XFD replacement
Use documented procedure in Direction 5364191:
Replacement / System Cabinet Replacement / XFA and XFD-PS
Replacement / XFD-PS
a. Perform the XFD-PS replacement
2. Alternative way to pull out the XFD
Notice: Only use this method in case XFD can’t be pulled out
according to the normal procedure
a. Perform XFD Power Panel Assy removal by using documented
procedure in Direction 5364191:
Replacement / System Cabinet Replacement / Cabinet Rear
Panel / XFD Power Panel Assy
b. Perform the XFD-PS replacement
Questions: 1. What is the use of SW1 in the front panel of XFD-PS?
__________________________________________________________
2. What will happen when power on XFD-PS if the SW1 isn’t turned
back to operation position?
GE Healthcare MR5047 SV Lab Guide
Page: 16 of 24
__________________________________________________________
3. What should be checked before pushing into the XFD-PS?
__________________________________________________________
4. What is the correct order of removing the terminal of power panel?
__________________________________________________________
5. What should
,be checked before power on the system?
__________________________________________________________
Review: Questions for instructor:
______________________________________________________
______________________________________________________
______________________________________________________
GE Healthcare MR5047 SV Lab Guide
Page: 17 of 24
9. Cabinet Monitor Lab
Description: This lab will familiarize the student with the functions of the Cabinet
Monitor and how to troubleshoot water leak in SV system.
Time Required: 1 hour
Learning
Objective(s):
Navigate the service documentation
Perform cabinet monitor functional check
Practice how to measure the sensor of water flow
Know how to isolate trouble with SF checker
Support Materials
Required:
Service Methods procedures 5364191 on disk, or as loaded onto
system computer. Refer to Cabinet Monitor functional check.
Tools Required: SF checker, DVM and Standard hand tools
Safety
Considerations:
Follow all LOTO procedures. They are posted in the lab and
incorporated in the documented procedures
Steps:
1. Cabinet Monitor Functional Check
Use documented procedure in Direction 5364191:
Troubleshooting / Troubleshooting Steering Guide / Water Leak
Troubleshooting / 1 Overview / Cabinet Monitor Functional
Check
a. Perform Cabinet Monitor Functional Diagnostic
b. Disconnect white connector on the rear panel of LCS or MCS,
then perform step a to check if any error happens
c. Restore the white connector
d. Measure the resistance between pin1 and pin4 of LCS or MCS
when they power on or power off
e. Disconnect connector J3 on the cabinet monitor. Observe what
will happen and record the error message
GE Healthcare MR5047 SV Lab Guide
Page: 18 of 24
f. Install the SF checker to J3 on the cabinet monitor. Observe
what will happen
g. Remove the SF checker and restore connector J3
h. Simulate leakage on the Leak Sensors3. Observe what will
happen and record the error message
i. Simulate leakage on the Leak Sensor2. Observe what will happen
and record the error message.
Questions:
1. Is there any difference of the diagnostic result between step 1.a
and step1.b? How to explain it?
__________________________________________________________
2. What kind of switch is the sensor of water flow?
__________________________________________________________
3. How the cabinet monitor identifies three status of leak sensor:
Disconnected, normal and leakage detected?
__________________________________________________________
Review:
Questions for instructor:
______________________________________________________
______________________________________________________
______________________________________________________
__________________________________________________________
GE Healthcare MR5047 SV Lab Guide
Page: 19 of 24
10. Fixed-Table Actuator Lab
Description: This lab will step the student through the procedure of replacing the
table actuator
Time Required: 2 Hours
Learning
Objective(s):
Navigate the service documentation
Remove and reinstall the table actuator
Support Materials
Required:
Service Methods procedures – 5364191 on disk, or as loaded onto
system computer. Refer to Replacement/Fix Table/Table Electrical
Actuator.
(http://3.28.123.26:8000/mr/cdrom/5364191-
2EN/root/data/Signa_EXCITE/content/1100957.htm)
Tools Required: Standard non-magnetic hand tools
Safety
Considerations:
Follow all LOTO procedures. They are posted in the lab and
incorporated in the documented procedures
Steps:
1. Table Electrical Actuator Replacement
Follow the documented procedure in Direction 5364191.
Warning: The table must be removed from the
magnet room for this procedure.
Continue to Fixed Table Adjustment Lab once the table has been
remounted to the magnet
Questions:
1. What method is used to align the mounting screw holes?
__________________________________________________
Review:
Questions for instructor:
______________________________________________________
______________________________________________________
______________________________________________________
GE Healthcare MR5047 SV Lab Guide
Page: 20 of 24
11. Fixed-Table Installation & Adjustment Lab
Description: This lab will familiarize the student with the adjustments needed to
align the table to the gantry and insure all safety/interlock
mechanisms are operational
Time Required: 2 Hour
Learning
Objective(s):
Navigate the service documentation
Dock Frame Installation
Level table and adjust top height
Check and adjust lock and latch adjustments
Check and adjust cradle emergency latch
Support Materials
Required:
Service Methods procedures – 5364191 on disk, or as loaded onto
system computer.
Tools Required: Standard non-magnetic hand tools
Safety
Considerations:
Follow all LOTO procedures. They are posted in the lab and
incorporated in the documented procedures
Steps: 1. Dock Frame Installation
Use documented procedure in Direction 5364191:
Installation / Guided Mechanical Install Flow for Optima MR360
/ Brivo MR355 / Illustration 2: 1st day / Dock Frame Assy
(http://3.28.123.26:8000/mr/cdrom/5364191-
2EN/root/data/Signa_EXCITE/content/1097184.htm)
a. Perform Dock Frame Installation Procedure
2. Top Height Adjustment
Use documented procedure in Direction 5364191:
Adjust/Cals/ Fixed Table / TOP HEIGHT ADJUSTMENT
a. Perform Procedure steps 1 - 7
3. Cradle Side Lock Adjustment
Use documented procedure in Direction 5364191:
Adjust/Cals/ Fixed Table / CRADLE SLIDE LOCK ADJUSTMENT
a. Perform Procedure steps 1 - 8
4. Cradle Guide Latch
Use documented procedure in Direction 5364191:
Adjust/Cals/ Fixed Table / CRADLE GUIDE RAIL LATCH
ADJUSTMENT
a. Perform Procedure 4.1, steps 1 - 4
5. Cradle Emergency Release
GE Healthcare MR5047 SV Lab Guide
Page: 21 of 24
Use documented procedure in Direction 5364191:
Functional Checks / Fixed Table / CRADLE EMERGENCY RELEASE
CHECK AND ADJUSTMENT
a. Perform Procedure 3.2 steps 1 - 9
6. Cradle Home Sensor
Use documented procedure in Direction 5364191:
Functional Cks / Fixed Table / CRADLE HOME SENSOR CHECK
a. Perform Procedure steps 1 - 8
Questions: 1. How do you know if the table height is adjusted properly?
__________________________________________________
2. What is the adjustment length of the delrin lock?
__________________________________________________
3. What is the maximum travel distance for the cradle emergency
lock release ramp”?
__________________________________________________
Review: Questions for instructor:
______________________________________________________
______________________________________________________
______________________________________________________
GE Healthcare MR5047 SV Lab Guide
Page: 22 of 24
12. Express Coil Lab
Description: This lab will familiarize the student with the setup and image quality
assurance tests for the Express coil.
Time Required: 1 Hour
Learning
Objective(s):
Navigate the service documentation
Position the phantoms and coils for testing
Perform the Multi-Coil Quality assurance Tool
Support Materials
Required:
Service Methods procedures 5364191 on disk, or as loaded onto
system computer. Refer to Express coil SNR test .
(http://3.28.123.26:8000/mr/cdrom/5364191-
2EN/root/data/Signa_EXCITE/content/1095326.htm)
Tools Required: Unified Phantom
Phantom positioners, pad, and strap
Safety
Considerations:
Follow all LOTO procedures. They are posted in the lab and
incorporated in the documented procedures
Steps: 1. Express Coil MCQA Test
Use documented procedure in Direction 5364191:
Adjust/Cals / Surface and Head Coils / 1.5T Coil Vendor
,Manuals
/ Table 11: Express Coil / SNR Test
a. Perform procedure 3.1 step 1 –3 to setup HNA coil and position
phantom
b. Perform step 4 to run MCQA tool to test mode HNA+PA
c. Perform procedure 3.1 step 1 –4 to setup AA coil and position
phantom
d. Perform step 5 to run MCQA tool to test mode AA+PA
e. Compare and record test results
Questions: 1. What coil segment would be indicated if there was a failure test 2,
Sig_Img 4 in mode HNA+PA ?
__________________________________________________
2. What coil segment would be indicated if there was a failure test 2,
Sig_Img 4 in mode AA+PA ?
__________________________________________________
Review: Questions for instructor:
______________________________________________________
______________________________________________________
______________________________________________________
GE Healthcare MR5047 SV Lab Guide
Page: 23 of 24
13. MCR III Tool Lab
Description: This lab will familiarize the student with the MCR3 tool hardware setup
for Port A and Express PA coil, and software usage.
Time Required: 1 Hour
Learning
Objective(s):
Navigate the service documentation
Set up the hardware connection for Port A and Express PA coil
Know how to run the software and read the test result
Support Materials
Required:
Service Methods procedures 5364191 on disk, or as loaded onto
system computer. Refer to MCR III Tool for MR355/360.
(http://3.28.123.26:8000/mr/cdrom/5364191-
2EN/root/data/Signa_EXCITE/content/1084956.htm)
Tools Required: MCR III Tool
Express PA Coil RF Cable Assy for MCR III Tool
Express PA Coil DC Cable Assy for MCR III Tool
Safety
Considerations:
Strong Magnetic Field! Ferrous materials can become dangerous
projectiles in the presence of the magnetic field Produced by the
Signa Magnet. Do not Bring any ferromagnetic tools or equipment
into the magnet room.
Steps: 1. MCR III Diagnostic
Use documented procedure in Direction 5364191:
Troubleshooting / Image Quality Troubleshooting Tools / MCR
III Tool for Optima MR360 / Brivo MR355
a. Perform Port A, Express PA coil test
Questions: 1. What is the function of the two-pin connector?
__________________________________________________
Review: Questions for instructor:
______________________________________________________
______________________________________________________
______________________________________________________
GE Healthcare MR5047 SV Lab Guide
Page: 24 of 24
14. Modified Calibration Tool Lab
Description:
This lab will familiarize the student with the modified calibration tool
in SV system, including LV Shim, DQA Tool II, Grafidy 3 and EPI white
pixel test.
Time Required: 2.5 hours
Learning
Objective(s):
Navigate the service documentation
Practice the modified calibration tools
Support Materials
Required:
Service Methods procedures 5364191 on disk, or as loaded onto
system computer.
Tools Required:
LV Shim Phantom Assembly
Nesting Plate Assembly
DQA III Phantom
Grafidy 3 Kit
Split Head Coil
Safety
Considerations:
1. Follow all LOTO procedures. They are posted in the lab and
incorporated in the documented procedures
2. Take care when moving LV Shim phantom.
Steps:
1. Use Modified Calibration Tool
Use documented procedure in Direction 5364191:
Installation / Install and Calibration Wizard for Signa Brivo
MR355/Optima MR360 / 4 Procedure / 4.4 Calibrations Guided
Flow
a. Perform a saveInfo to save the current calibration
b. Perform LV Shim
c. Perform DQA Tool II
d. Perform Grafidy 3
e. Perform EPI White Pixel test
f. Restore the saveInfo of the system
Review:
Questions for instructor:
______________________________________________________
______________________________________________________
______________________________________________________
The information in this course is
FOR TRAINING PURPOSES ONLY
and not to be used as
promotional material. The
information may not be current
or appropriate for all systems.
Before working on any
equipment consult appropriate
current service documentation.
Failure to follow procedures in
current service documentation
or misuse of the course
information may result in
equipment damage, personal
injury or death.
The primary sources of the
material contained in this
course are released Service,
Sales, or Marketing
documentation.
GE Healthcare
MyWorkshop Book Number: 51232
SV Full Service Component
Identification Activity
Page 2
System Cabinet
1.
2.
3.
4.
5.
6.
2
3
4
5
6
1
Page 3
System Cabinet
1.
2.
3.
4.
5.
6.
1
3
4
2
5
6
Page 4
System Cabinet
1.
2.
3.
4.
5.
6.
4
5
6
2
3
1
Page 5
System Cabinet
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
2
3
4
5
6
7
8
9
10
11
1
Page 6
Magnet Side Kit
1.
2.
3. 1
2
3
Page 7
Rear Pedestal
1.
2.
3.
4.
5.
6.
7.
4
3
1
2
56
7
Page 8
LPCA
1.
2. 1
2
Page 9
OC Computer
1.
2.
3.
4.
5.
6.
2
3
4
5
6
1
,board and Mega Switch by Convert Board located under
the foot pedal cover, which one cable is for receiving RF signal from PA coil and Multicoil
bias, and another cable is for providing DC power for PA coil.
Because it is a little difficult and time consuming to disassemble the cradle, and handling PA
coil needs great care, for easy, safe and quick reason, the cradle and built-in PA coil
composed one FRU. Two person need when replacement because the weight is more than
16KG.
27
The trough on the newly designed Front Bridge is used to holding the cable track on the
bottom of the cradle.
28
For the reason of cost down and simplication, some coils use the unified phantom to
replace their own phantom. Refer to service manual for detail procedure of SNR test.
Unified phantom also can be used in HDe, HDx after the corresponding service pack is
installed.
29
• In all GE MR systems prior to SV, HDe and HDx, the keyboard and mouse communicated
via PS2 port.
• The SV system uses an optical mouse and both the keyboard and mouse communicate
with the host via USB.
• As the mouse, keyboard and monitor are all black, we can't leave the SCIM behind!
• Nothing about the SCIM's has changed except for the color.
• The biggest change you'll see at the operator workspace is the reduction from two
screens to one larger, wide-screen monitor.
• The user interface has been updated so that all information from both screens is now on
the widescreen.
30
There is only one system cabinet that covers the function of RFS, HFD cabinet and part of
penetration panel.
The following two slides show the inner components under the covers.
Note:
1) Store L/R cover carefully to prevent the distortion
2) Take care of G-Filter and power panel when moving the cabinet
31
32
There are three access window on the back of the system cabinet.
1. Access window for back of CAM Lite2 and GP3, ICN are used during installation.
2. Access window for PDU FRU only be used during replacement
33
The gradient system in SV is called XFD. It contains three gradient amplifiers XFA and one
gradient power supply XFD-PS. They are all water-cooled. A leak sensor is installed in front of
XFD to detect leakage of water pipe connectors. XFD slew rate is 100, which is twice as
much as HFD-lite used in HDe system.
34
In order to improve gradient system service capability, traditional cables and connectors are
not used for connection between XFAs and XFD-PS in SV system. XFD-PS power output and
XFA power input are used plugs and sockets, which is connected by a newly designed
component, Power panel. And XFA gradient output are also used plugs, which is directly
connected to G-Filter. With this kind of structure, it is more convenient to swap XFA than any
previous system when troubleshooting.
35
Power panel connects XFAs and XFD-PS. It can be accessed by removing its cover in the magnet room. If
the power panel needs to be replaced, be assure to wait for at least 15 minutes after XFD power-off. And
the nethermost plug must be removed before removing other plugs.
36
37
38
There are 108 screw holes on the frame connected to the system cabinet. But only 60
screws are supplied. Fix the System Cabinet with washer and screws at every two hole pitch.
Do not install screw in every hole.
39
40
The three phase alternating current for MCS and 8KW LCS must have the correct phase
position to operate. If the phase rotation is incorrect, there is no indication of power to the
system. Swap two phase leads, observing LOTO precautions. The 4KW LCS doesn’t have
this requirement. On the top of system cabinet, there are only one terminal connected to
MCS or 8KW LCS so if the configuration has another 4KW LCS, it will get AC power from the
8KW LCS.
41
42
43
44
• There are five kinds of water cooling system configurations.
• Type A, B and C have their individual plumbing assembly.
• In type B and type E, Facility water is provided by the customer. The water should meet
the specification in the pre installation manual. In not, it is possible to damage our
equipment.
• In Type D, the BRM chiller and Air-cooled cryogen compressor are only available to 200V
AC power input. So the facility power can only be 200V AC.
• Type E is only for EMEA. A 3 Phase AC stepdown transformer should be installed which
supplies the 200V to BRM chiller.
45
46
47
48
49
50
51
52
53
54
55
In previous installation of outdoor water chillers, the vendor used different valves for
connection. The quality can’t be assured by this way. Now the Plumbing Assy will be
shipped to site according to the different configuration of cooling system.
56
In order to meet the market requirement, SV system contains two kinds of configuration:
Optima MR360 and Brivo MR355.
For the customers who are familiar with GE MR system, Optima MR360 can provide
affordable, versatile MR scanning to meet everyday patient needs with
uncompromised(uncompromising) imaging, workflow, service and support.
For the customers who are new to MR system, they will learn to use Brivo MR355 without
too much training. Function of Simple UI makes scan operation much easier than any
previous MR.
Note: OIC = Outpatient Imaging Center
57
In general, Optima MR360 & Brivo MR355 have the same basic hardware configuration.
Fixed Table with Express coil can be used in both Optima MR360 and Brivo MR355.
Detachable Table can only be used in Optima MR360.
58
• Optima MR360 and Brivo MR355 have the same application software with different
software options and coil options.
• The software options available in MR360 are more than that in MR355, which means
MR360 is more powerful than MR355.
• Optima MR360 can use all kinds of coils. Brivo MR355 has four kinds of configurations .
Only the coils included in the configuration can be used, which is controlled by the Option
key.
• There is no CoilConfig File in SV, which is replaced by CoilDB. CoilDB is a predefined Coil
Database, which stores the config information of all coils. Users can’t add, delete and
modify coil config information in CoilDB, but can view the content by Coil Database
Explorer in Service Browser.
59
• For the purpose of market sales, Brivo 355 has four kinds of configuration: Main, Main +
Knee, Main + Shoulder and Premium.
• In Brivo Main configuration, only 8 coils listed in the left table can be used.
• In Brivo Main + Knee configuration, not only the initial 8 coils but also Quad Extremity coil
can be used.
• In Brivo Main + Shoulder configuration, the 8 initial coils and 3 channel shoulder coil can
be used.
• In Brivo Premium configuration, 10 coils can be used, including the Quad extremity, 3
channel shoulder coil and the 8 coils in Main configuration.
• So, only the coils included in the configuration can be used. For example, 3Ch Shoulder
coil can’t be used in Brivo Main and Brivo Main + Knee configurations. Be sure to confirm
what kind of configuration is in your site before using any coils, otherwise the coil may be
not identified.
• There is no this limitation in Optima MR360. But MR360 has to follow another limitation
according to different table type.(Refer to the next page)
60
• HDe can be upgraded to SV due to higher performance than HDe. Almost everything will
be replaced except the LCC magnet.
• But only type B of HDe with equipment room configuration can be upgraded to Optima
MR360 system.
• For type B of HDe with water cooled shield cooler, shield cooler must be removed and will
be replaced by F-50.
• For type B of HDe with air cooled shield cooler, shield cooler will be reused. No need to
remove.
• Optima MR360 configurations upgraded from HDe are Type C and Type D with detachable
table.
• For detailed information, refer to Signa HDe to Optima MR360 System Upgrade Manual
(P/N: 5410639-1EN Rev 1.1)
61
Brivo MR355 can be upgraded to Optima MR360 after
,uninstalling the previous one of Brivo
option keys and installing MR360 Option Key. No hardware change is needed.
62
63
64
65
66
67
With Favorite Protocol, users can select their favorite protocols at hand graphically instead
of opening protocol selector window. It has two kinds of favorite protocols:
• Ready Favorite Protocols, also called Ready Shortcuts
• Custom Favorite Protocols, also called custom shortcuts
Ready Favorite Protocols are predefined and cannot be changed by users. User can define
Custom Favorite Protocols from any protocols.
68
69
Simple UI is designed for inexperienced customers who can't and those who don't want to
handle complicated MR physical parameters as “novice mode”. It allows switching to
“Advanced” and “Expert” UI modes at any time (except after the task is saved). The
inexperienced users can finish the routing scan in Simple UI.
Basic Scan Mode is designed for the routine examination. Most of all scan parameters are
hidden to simplify the visual look and avoid changing the parameter without careful
management on the image quality. The Slider Bar is the way to select different sets of the
scan parameters.
70
The Advanced Scan Mode and Expert Scan Mode is to control all scan parameters in detail.
The scan mode can be switched when the [Mode change] button is selected.
71
Default Scan Screen mode can be set as System Preference.
The default Scan Screen mode applied at the beginning of the exam.
Note: Scan Screen mode is kept in a exam until the mode is changed.
72
1. Tradeoff is shown on the slider bar.
Basically the SNR is kept for all positions, but the spatial resolution is different.
2. Following scan parameters are propagated to the other position of the Slider Bar.
Scan plane, Start/End location, Auto shim, Tracker length/thickness, Patient entry/position,
Coil configuration, Series description, Contrast information
3. Slider Bar protocols are not allowed to save and duplicate.
Proto Copy will be helpful to make the protocol from images scanned with the Slider Bar
protocol. The protocol generated by Proto Copy includes just one scan parameter set of the
Slider Bar protocol.
4. Slider Bar function doesn’t apply to all the protocols. Only the protocols in the above list
are OK to use. The protocols selected from the Ready Shortcuts are Ok to use too.
A easy way to check:
If there are short vertical lines showed on the Slider Bar, it means the protocol can use this
function.
73
Protocol Selector are totally different from the previous version. There are the specific user
interface for Adult and Pediatric protocols.
74
Protocol Template only can be accessed in GE protocol library. In this template, all the
common used protocol can be found.
75
Image can be previewed in this new version of Image Management.
76
77
78
79
80
You can access the same diagnostic items by System Function or Hardware Location. Their
functions are same.
81
IP Protection is new designed Software Tool for security improved of GE proprietary tool.
The function is optional that can be closed as defined by local service organization. When IP
Protection is enabled, the field service personnel will be required to input an authorization
code to access the service browser to utilize GE proprietary tools. The authorization code can
be generated automatically in the back-office sever after one IPP Key is input that is
generated by MR host software.
NOTICE: Service IP Disable Key (M50002DJ) shall be ordered to disable the function.
NOTCE: IP protection is just available in China as pilot prior to M4. The design will be improved
based on the field feedback. The global deployment plan will be finalized prior to M4.
Main function consists of:
• Optional (Default is active, “iprotection” option disable it)
• Only lock Class C/M tools( USB service key shall be presented)
• Auto-run( User generate the activation key by themselves)
• B/S structure( web browser to log in server, like e-License)
• Customized useful-life ( period of validity from 48 hrs to 30 days, only for administrator
account)
• User management and Event log (Add/Delete user account, log activation code generation
time, user, hospital…, only for administrator account)
• Manually generate in emergency ( Server down, Administrator can generate key by
another SW)
82
83
The E-Reporting Tool is a web-based application that allows for better communication
between the customer and the GE service engineer. Communication is one-way with this tool
from the service engineer to the customer. When a Field Engineer (FE) or Online Engineer
(OLE) provides service at a customer site, the E-Reporting Tool is used to document issues
found or service performed. It is also used to provide recommendations for customer follow-
up, for example, suggesting that the customer improve room temperature. After the service
engineer has created a service report, the customer can view and print the report at any
time.
There are three types of reports in the E-Reporting Tool:
1) Emergency and Routine Service Report - (The Field Engineer primarily uses this page to
communicate to the customer any information from emergency or routine service calls.) This
report is created, for example, when an FE or OLE is called onsite to resolve an emergency
site issue and wants to maintain a log of the discussion that occurred between site
personnel and the engineer. The service engineer is also able to communicate with the
customer about work that was done after hours at that site.
2) Planned Maintenance Service Report - (The Field Engineer primarily uses this page to
communicate to the customer any PM information from planned maintenance service calls.)
An FE or OLE may create this report to communicate to site personnel about the details of
planned maintenance that was performed onsite by GE. The report could also contain any
site-related issues that the customer should correct or that GE needs to fix.
3) Predictive Maintenance Service Report - (The OnLine Center primarily uses this page to
communicate to the customer any information from predictive maintenance service calls.)
This report may be created by an FE or OLE when performing proactive work to remedy a
problem that will occur onsite if site conditions are not corrected. This may involve site-
related issues that could impact product performance. Examples include temperature and
humidity that is out of specification, or MCQA results that indicate a problem with site coils.
84
In SV systems, it is easy to enable or disable autologin function. If the current mode is
unclear, use the STATUS function to check.
85
86
87
88
89
90
CAN stands for Controller Area Network, which is a vehicle bus standard designed to allow
microcontrollers and devices to communicate with each other within a vehicle without a
host computer. It is also used in other areas such as industrial automation and medical
equipment. Since MGD was introduced into GE MR system, CAN replaces the MDS link
gradually. In different systems, the components connected through CAN link are different.
1. SCP3 is in charge of receiving and managing the system data, sending control instructions
and monitoring the status of all connected node devices through the CAN link.
2. Every node device has its own built-in CAN communication controller. All controllers are
powered by a single isolated +24V power supply in the cabinet monitor.
3. CAN needs terminator resistor to balance the impendence of the network. One 120ohms
resistor is embedded in the SCP3 board, and another 120ohms resistor is installed on the
farthest node - Cabinet Monitor.
4. The DC24V ISO power is provided by Driver Module in HD, HDx system. In Emergency Off
mode, Driver Module will lose power and the CAN link is broken. The system requires a TPS
reset after power restoration. In SV system however,
,CAN 24V DC power is supplied by
Cabinet Monitor, and it is still energized when system is in Emergency Off mode, so TPS reset
is unnecessary after power is restored.
5. In service browser, CAN diagnostic tool can be used to check if the CAN link works
normally. Cabinet Monitor should be always connected because it supplies the 24V DC
power for the whole CAN link. We can bypass one node by removing the CAN in and out
cables and directly connecting the two cables except the driver module lite, because the
CAN cables connectors in driver module are both male.
91
This graphic displays the Ethernet routes and serial service for the SV system.
• Because CAM lite2 has no APS, so port 1 in Term Server is not used any more.
• Image data, status and control signal are transferred by the 1000M Ethernet switch.
• A VERY IMPORTANT THING TO REMEMBER: Image Compute Nodes are computers like your
PC… they must be shut down properly or they risk operating system corruption.
• ALWAYS follow service procedures when removing power to ICNs – shut them down via a
“soft power off” at the GOC, not by pushing the power button on the ICNs themselves or
by pulling the power cord. We will say this many times in this training!! In addition,
removing power to the Systems cabinet (or RF/Systems cabinet) without properly shutting
down the Volume Recon Engine subsystem will induce the possibility of corrupting the
operating system on the ICNs!!
• VRE can be safely powered down by 1) shutting down the host computer or 2) Using the
VRE power utility from the common service desktop
92
The Common MR Communications Link is first used in the DVMR product development. In SV,
we use the same communication link.
SerialLite is a serial protocol developed by Altera Corp. and Innocor Ltd. Altera provides an IP
core implementation of SerialLite that utilizes the high-speed serial transceivers in their
Stratix GX FPGAs. This communications link assumes implementation in a Stratix GX FPGA
using the Altera IP core.
The DVMR Common Communications Link, or dvmrlink, defines bidirectional data transfer
within the DVMR system. This includes data transfer between the CAM chassis and remote
devices, including RF Exciters, RF Receivers, and data transfer between an RF Receiver and a
VRF installed a VRE. IRF3 board is the interface of DVMR link.
1. DATA TRANSFER FROM CAM TO REMOTE DEVICES
Data transfer from CAM to the remote devices includes both synchronous and
asynchronous data. The Sequence Bus, which contains SSP (Sequence Scan Protocol)
Commands, RF data, and gradient data, is the source of synchronous data. The CAM CPUs,
via the PCI bus, are the sources of asynchronous data. Asynchronous data transfer
effectively provides PCI write access to the remote devices. Synchronous and asynchronous
data is merged into an CAM data packet.
2. DATA TRANSFER FROM REMOTE DEVICES
Data transfer from the remote devices include status information, sample data, echoed SSP
commands, and echoed PCI commands.
3. Data Transfer from VRF to Remote Receivers
In SV system, the 8 channel RF signals are digitized by ADC in RRX and converted into optical
signal, send to VRF in VRE via VRF Data Link.
93
This block diagram shows RF receive chain in SV system. If the system configuration is
MR360 without Express coil, J22 in Mega Switch has no cable connection.
Mega Switch selects different input signal from Port A coil, PA coil and body coil, and
amplifies the selected signal, and converts them to intermediate frequency signal, which
center frequency is about 16MHz.
Then, RRX gets the eight channel RF signals and digitizes them with ADC, and converts them
to optical signal, and send to VRF in VRE via VRF Data Link.
VRF converts the optical signal to electrical signal and filters them, then sends them to ICN
for image reconstruction.
Then the reconstructed image is send to host via Ethernet switch.
94
95
96
During RF transmitting, Unblank signal will be generated and sent to synchronize
remote devices. Digital Unblank signal is sent from CAM lite to Exciter via DVMR link.
The analog Unblank signal is generated by the Exciter.
Then, the analog Unblank signal is sent to Driver module lite Unblank distributor,
which outputs same Unblank signals and sends to UPM1, UPM2, RF amplifier interface
board and SRFD3.
On the other hand, Unblank and AC Power Sense signal will be sent to the Magnet
Monitor.
1. Unblank signal may aid in executing ADC algorithms for improving accuracy of the
sensed ADC inputs that have spikes due to RF noise coupling.
2. Power Sense indicates whether or not the System Cabinet is powered-up. The
resulting state shall be made available for the SBC (Single Board Computer) in Magnet
Monitor to read. This will allow the determination of the validity of the Unblank signal.
97
PCIAA is functionally similar to GOCAA (Global Operator Console Audio Assembly), but
is only used for audio communication between operator and patient. It locates in the
host PC and is mounted on one PCI socket. But PCIAA has no any electrical connection
with PCI socket, only occupies one PCI slot to fix.
The cable (E3047) between SCIM and PCIAA are new designed part. The mouse are
keyboard are all USB port.
98
Simple OC consists of host PC and chassis.
Host, LCD and Ethernet switch get the power supply from the power strip in the
chassis. There is a circuit breaker on the power strip, which can only be accessed after
right cover removal.
Pay attention to the service key after inserted in the front USB port. It is easy to be
broken when anyone passes by the OC! It is suggested that the service key had better
to be inserted in the rear USB port.
The Option keys are bounded to the ethernet card built in the main board. But the
main board is not a FRU. The whole PC has to be replaced if the main board doesn’t
work, and all the software options has to be regenerated in the eLicense website.
The following is the FRU List in Simple OC:
Host PC ( THTF M10)
Memory ( 2GB DDR2 677 DIMM)
DVD RW ( PLEXTOR PX-806SA)
HDD ( Western Digital WD3202ABYS-0)
Dust Filter
LAN Switch
PCIAA Collector
OC Chassis Collector
Cable Collector
99
The input power for THTF host should be 100~240VAC.
SCSI tower isn’t used in SV system any more. Only one DVD-RW driver is installed in the
host. LFC, save and restore info are all executed by this driver. It is fixed by two screws
on the left side. There are no fixed screws on the other side. It is easy to be pulled out
with a little of force after removing the screws.
The glue on the bracket of the memory socket is to assure that the memory will not
become loose during transportation. Be careful to remove the glue when replacing the
memory.
100
This slide shows the detailed configuration of THTF host.
101
• PCIAA (PCI Audio Assembly) Kit contains one ribbon cable and two boards: PCIAA
board and PCIAA Extended board.
• There are several audio adjustment potentiometers on the PCIAA extended board
for adjusting patient and operator voice levels. The Minimum Patient Speaker
Volume is fixed, there is no adjustment for it .
• It is easy to forget to connect the power cable to the PCIAA. Be sure that it’s been
connected before you close the host cover. Otherwise, TPS reset will not respond.
102
103
104
105
• Due to 2.15 Farad super capacitor used in gradient power supply, the rated power
of system cabinet PDU is decreased to 25KW.
• 3 phase lines and 1 ground wire are connected between MDP (Main Disconnect
Panels) and PDU.
• Input voltage is selected by inserting the incoming power cables into the
corresponding transformer terminal.
• DIP switches, under the cover, set the overload and short circuit trip on input power
as on other systems.
• There are two cables connected to the PDU control board. E-Stop control cable are
connected to three emergency off button. Another cable is from cabinet monitor,
,which trip E-off contactor when issues are detected, and switch on or off Nigh
mode contactor.
• When the PDU is initially installed, all breakers are off. Be sure to observe the
correct order of switching on the PDU breakers to ensure coolant is flowing before
heat generating components, such RF amplifier, gradient amplifiers and gradient
power supply, are powered up.
106
The 25KVA is a new design part. The power capacity is less than former products
because the gradient power supply uses the super capacitors to store power energy in
SV system.
Control signals to Main Breaker:
1. Power Off Button Control: This signal is from Power Off button. Main breaker will trip
if press this button.
2. Transformer Temperature Sensor Control: This signal is from 3 phase transformer in
PDU. Main breaker will trip if the temperature of the transformer is over the limit .
3. Main Breaker Trip Control: Refer to 2.c in this page.
Control signals to Control Board:
1. Input control signal:
a. EMO Reset Button Control: This signal is from EMO Reset Button. E-off contactor
and Night Mode contactor will be energized after pressed this button.
b. E-Stop Control: This signal is from outside E-Stop circuit (Refer to the next page). E-
off contactor and Night Mode contactor will lost power when pressed any E-Stop
button.
c. Cabinet Monitor Control: This signal is from cabinet monitor. It contains two kinds
of signal. One is for controlling E-Off contactor, the other is for controlling Night Mode
contactor. For detailed information. Refer to Cabinet Monitor section in this
documentation.
107
2. Output control signal:
a. E-Off Contactor Trip Control
b. Night Mode Contactor Trip Control
c. Main Breaker Trip Control: In HDe system, when leak sensor 1 detects any leakage,
cabinet monitor will send control signal to PDU to trip the Main Breaker. The whole
system will lost power input. But in SV system, leak sensor 1 is removed. So this signal
is not used in SV any more although the control circuit is still there.
108
MPS (Magnet Power Supply) is for ramping up power supply. It will have no output
after E-stop contactor is tripped.
109
There are five Switching AC to DC power supplies located on the upper right of the
system cabinet.
DCPS for DM_Lite, Exciter, Mega SW & RRX only have input power 120VAC, no control
signal, which means that the three DCPS have stable and uncontrolled output once
the input power 120VAC is engaged.
Output of DCPS for DM lite: +24V/+15.38/-15v/+12V/-12V/+10V/-10V
Output of DCPS for Exciter: +5.3V/+12.6V/+15.8V/+7.4V/-7.4V
Output of DCPS for Mega SW & RRx: -9.25V/+9.75V/+6.625V
DCPS LED Power includes the following power and signal:
1. Provide +24V, +15.25V, -15.25V, 5V_PAL(For Patient Alignment Laser) and 5V_BL
(For Bore Light) to LED Power Box
2. Output Mag_DC_OK status signal to PHPS Lite2
3. Receive control signals from PHPS Lite2, including Remote On/Off signal of BL and
PAL.
DCPS PHPS-Lite2 includes the following power and signal:
1. Provide +42V and +8V to PHPS-Lite2,
2. Provide +42V to DOCK
3. Provide 120 VAC to PHPS-Lite2 for controlling Patient Blower
4. Output Table_DC_OK status signal to PHPS Lite2
5. Receive +42V sensing signal from the dock motor to adjust the output +42V to Dock
motor.
110
111
1.Overview
The PHPS stands for Patient Handling Power Supply. It functionally replaces the SSM
(System Support Module). There are three types of PHPS for different type of MR. The
above table shows the difference between PHPS, PHPS-Lite and PHPS-Lite2.
2. Basic principle
The PHPS-Lite2 is mounted in the Cabinet.
It consists of a PHPS Control board, a CAN Communication Core (CCC) Board which is
physically mounted as a daughter board on the control board and a pulse width
modulated Servo amplifier.
DC +42V and +8V are supplied from the outside Power Supply for PHPS-Lite2.
This power in turn is used to generate regulated +5V and +3.3V locally for use on the
control board and +42V for Servo Amplifier.
112
1) The Control Board Assembly gets command signal serially from IRF I/O J23 in CAM-Lite2 cabinet. The
Control Board generates signals needed for control of the following entities:
• The Patient Alignment Laser
• The Bore Light
• The Patient comfort fan relay
• The enable signal needed by the Servo amplifier
• The output relays in the longitudinal motor drive circuit
• Current feedback signal for monitoring the dock motor current
• Communicates status of power supplies and servo amplifier to the CCC board
2) The Servo Amplifier is a Pulse Width Modulated (PWM) device used for powering the longitudinal
motor. DC 42V generated by the control board is the base input voltage. The SRI send command signal
of In and Out or Fast In and Fast Out directly to Servo Amplifier to drive the longitudinal motor.
3) The CAN Core Communication (CCC) board monitors the health of the power supplies and the status
of the servo amplifier, and communicates this information to the host. The CCC board receives power
supply voltages, digital I/O signals, and analog inputs corresponding to the power supply voltages
from the Control Board. The digital I/O of the CCC operates at 3.3V so all logic levels must be converted
to this level on the Control Board. Additionally, the CCC gets an isolated 24V from an external source
for powering the serial interface communications.
3. Block diagram
Signal Table_DC_OK comes from Power Supply for PHPS Lite2. PS for PHPS Lite2 provides DC +42V and
+8V to PHPS Lite2 by J6. The signal is activated when DC +42V is OK.
Signal Mag_DC_OK comes from Power Supply for LED Power Box. PS for LED Power Box provides +15V,
-15V and +24V DC power to LED Power Box and controlled DC +5V to Patient Alignment Laser and Bore
Light. The signal is activated when Power Supply works normally.
113
CAM Lite is short for Consolidated ASC MGD Lite chassis. It consolidates the
components of the Multi-Generational Data acquisition chassis (MGD) and Amplifier
Support Controller (ASC) chassis into one unit on HDe forward production Systems
cabinets. The CAM-Lite chassis combines the individual ASC and MGD boards into a
common chassis. There is no change in functionality.
114
115
116
The CAM-LITE2 is a consolidation of two subsystems, the MGD LITE and ASC LITE. It
contains a custom 6U cPCI midplane that is comprised of a single custom bus
segment, two cPCI bus segments and a common power system. The single custom
bus segment house the Universal Power Monitor (UPM) circuit card that provide RF
amplifier power monitoring function with redundancy and the Amplifier Interface (AIF)
circuit cards for the Narrow Band (NB). The functions within this custom bus segment
communicate to the MR system, utilizing CAN via the midplane. A single board
computer (AGP2) serves as the cPCI bus segment masters. A third processor board
(SCP3) provides serial interface functions and house a CAN mezzanine card. The GEHC
custom boards IRF I/O, SRF/TRF and STIF interface to one of the cPCI bus segments. A
custom, non-transparent, cPCI bridge board allow cPCI communication between the
primary and secondary cPCI midplane segments.
117
AGP(Applications Gateway Processor) or the AGP2 include a cPCI(Compact PCI) single
board computer. The AGP consists of a Motorola MCP750 or equivalent, and include a
64MB memory mezzanine module. The AGP-II consists of a Motorola MCP820 or
equivalent, and include a 128MB memory mezzanine module.
So AGP2 Board (Slot 6 F) is also sometimes referred to as MCP820. That is why the
terminal window shows 820 when we reset TPS while logging in AGP to monitor it
status using command mgd_term. AGP coordinates the work of the IRF, IRF I/O,
SRF/TRF, and STIF Boards to create the digital RF and gradient data needed to
generate the scan waveforms. The top cable, equipped with the male RJ-45 connector,
provides an Ethernet link between the Host and the AGP. During a TPS Reset,
,PSD data
download from Host and data transfers to the AGP via this link. The bottom cable,
equipped with the male RJ-45 connector, provides a service RS-232 serial link between
the AGP and the Host via the Term Server.
118
The SCP3 Board, also is called CPV3, is functionally update of a standard SCP and SCP2 Board, which
its performance increases dramatically. It is required for certain purchased scan options.
The following is the functional and performance enhancements over the CPV-3060 (SCPII) assembly:
• Compute performance: the SCPIII processor operates with a core clock of 667 MHz, compared to the
50MHz processor clock on the CPV-3060 processor
• Memory bus performance: the SCPIII processor hosts DDR2 SDRAM components on a dedicated
167MHz bus, with a raw bus transfer capacity of ~21 Gb/s, as compared to the 2.1 Gb/s raw bus
transfer capacity of the 66 MHz SDRAM components on the CPV-3060
• Memory array capacity: the SCPIII processor offers 64–512 Mbyte capacity at the main memory
array, compared to the 64–128 Mbyte capacity offered on the CPV-3060. The RAM configuration for GE
part is 256M.
• Front panel Ethernet port: the SCPIII hosts a 10/100/1000BaseT port as compared to the
10/100BaseTX Ethernet port on the CPV-3060
• CANBus interface ports: SCPIII provides two front panel CANBus 2.0 interfaces on the main board in
place of the ESD Electronics PMC331-1 CANBus controller module provided with the CPV-3060.
All front panel and primary backplane I/O functions of the SCP2 are maintained on the new board
except for the rear-access SMC port, which is not supported on the newer PowerQUICC processors.
SCP3 has one PMC site and provides two on-board CANBus 2.0 interface controllers in place of the
second PMC site provided on the SCP2. The SCP3’s internal hardware architecture is implemented
around a new MPC8360 PowerQUICC II Pro processor with a 667 MHz core clock.
The middle white cable, equipped with the male RJ-45 connector, provides an Ethernet link between
the Host and the SCP. During a TPS Reset data is transferred to the SCP via this link.
The bottom cable, equipped with an RJ-45 connector, provides a service RS-232 serial link between the
SCP and the Host via the Term Server.
119
The IRF3 is the third generation Interface and Remote RF Functions circuit board. It is designed for use
within the DVMR system and, despite many architectural similarities, maintains no backwards
compatibility with IRF or IRF2.
The IRF3 primarily serves as a communications hub within the CAM chassis, supporting
communication with one or two Remote Receiver devices (RRx), one or two Remote Exciter devices
(DTx), and the XGD gradient subsystem via five identical high-speed (2Gbit) serial fiber optic interfaces
implementing the DVMR Common Communications Link.
The IRF3 accepts an 80MHz reference clock input via a fiber optic receiver. This clock is used as the
master clock, both on-board and for the CAM system. CAM clocks, except PCI, are normally derived
from this reference. The FPGA provides logic to detect the absence of transitions on the reference
clock input. The FPGA also provides the ability to operate the CAM system stand-alone, using the PCI
clock instead of the reference clock, and, in fact, this is the default clock mode of the IRF3. Note that
system operation is asynchronous in this mode.
The IRF3 includes a front panel reset push button for CAM chassis reset. The debounced switch drives
the backplane reset request to the backplane system slots with a pulse of at least 100ms to initiate a
PCI reset.
The IRF3 interfaces to the CAM system through the backplane. These interfaces include the 32-bit 33
MHz PCI interface and the custom Sequence Bus. This also includes the CAM chassis temperature
monitoring portion of the SMC interface.
The IRF3 provides the same SMC inputs and outputs as IRF1 and IRF2. In SV, only Scan Room Door is
required.
The IRF3 provides five instances of the DVMR Common Communications Link that are identical and
interchangeable. The fiber optic transceiver modules have internal real-time diagnostic capability, and
measure internal module temperature, internal supply voltage, transmit bias current, transmit optical
output power and receive optical input power. Via the FPGA, these values are collected and made
available to the system via PCI.
Eight indicator LEDs on the front panel to provide general status information and to provide the status
of the communications links.
120
The RF Amplifier Support Chassis (ASC) will house two complete Universal Power Monitors for
redundancy. A UPM will consist of a UPB and up to two UPM RF Detector Boards (RFD) connected on
the ASC Backplane. Each UPM will be capable of monitoring forward and reflected RF power in a MR
system for 0.35T to 3T range of field strengths. In 1.5T SV system, only forward RF power is monitored.
The RF Detector Board is the front end of the UPM. They accept coupled RF power from the various
transmitter pathways in the MR systems, and execute Analog to Digital conversion (ADC) under the
control of UPM Processor Board (UPB), and convert them to a digitized power envelope.
The digitized power envelope is passed along to the UPB to be processed through mathematical
algorithms for the monitoring of RF power levels and for the detection of excessive power conditions.
The UPB will communicate status and fault/error conditions to the system through CAN as well as
supporting inter-module communications as necessary.
In SV and HDe system, RF AMP Interface board (AMPIF) is introduced to support SRFD3. The AMPIF
receives NB UNBLANK signal from the Driver Module and passes these UNBLANK signal to the RFPA.
When RF Output Power is above the limitation, UPB will create RF LOCK signal and send it to AMPIF.
AMPIF will send RF LOCK signal to RFPA to stop power output. And UPB sends inhibit Unblank signal to
the Drive Module to stop output the Unblank signal.
Attention: The AMPIF for CAM Lite is basically compatible in function with the AMPIF in ASC Lite of HDe.
But the pin assignment of RF LOCK on backboard is different between ASC LITE and CAM Lite. They
would be compatible by switching the pin assignment by switch on the board.
Note: NB = Narrow Band, BB = Broad Band, CW = Continuous Wave, BB and CW Unblank signals are
only for Multi-Nuclear Spectroscopy (MNS) amplifiers.
121
Driver Module Input and Output Signals:
1. Head T/R Bias (J6) to SRFD3: +8VDC(@4.25A)/-14.4VDC(@400mA)
2. Body T/R Bias (J7) to SRFD3: +5VDC(@5.5A)/-14.4VDC(@400mA)
3. Dynamic Disable Bias (J21/22/24/25) to Body Coil: 500VDC/ -1A(@-3.7VDC)
4. Direct Drive Bias (J23) to Body Hybrid: 500VDC/-1A(@-3.7VDC)
5. MC Aux Power (J3): +/-15VDC used by coils to generate preamp bias for preamps in
coils.
6. MC Preamp Power (J3): +15VDC used in LPCA to generate preamp bias for preamps
in LPCA.
7. MC Switch Control (J3): signals go to 8-Channel Switch Board to control 8-Channel
Switch Board operation.
8. MCD 0-7 (J5): Multicoil T/R Bias to multicoils (+3/+5/+7/-5VDC@(0.5A)).
9. Unblank signal input (J19) from Exciter goes to CAM Lite2 (J11, J12, J13). DM can
disable RF by disabling Unblank if bias faults.
10. CAN link (J1/J2): come from PHPS Lite2, go to GP3 from DM.
11. HV/TR/DD/MC Fault Disable Switches: service facilities in DM.
12. Serial High-speed data input (J4): This input will provide the high-speed differential
data necessary for “on-the-fly” coil switching. This data will be used only for intra-scan
modification of MC_SELECT to change the selection of the MC driver outputs.
It will modify only the drivers selected to switch to “receive mode” and will not have
the capability of modifying the driver’s positive output voltage level or any fault
detection levels.
122
123
RF Exciter Outputs
1. Analog RF excitation signal
2. 80MHZ Clock signal to VRF, IRF3, Mega Switch and RRX via Mega Switch
3. Local Oscillator (LO) signal and
,Loopback for Mega Switch
4. Unblank signal to Driver Module and magnet monitor
124
The VRE(Volume Reconstruction Engine ) replaces the Array Processor in the Signa
HDx, HDe and SV systems. It does the image processing and reconstruction functions
that were previously done by the Reflex AP's in the Excite2A and Excite HD systems.
The VRE in the SV system has only one ICN (Image Compute Node). The ICN is a Linux
PC in itself. One ICN is roughly equivalent to the processing power of a Reflex 800 AP
in the Excite HD. Whenever the system and/or MGD IP addresses are changed, the
VRE must reconfigure or it will not communicate with the system.
Sun 4100 and Sun 4170 can all be used in SV system. But because the Sun 4100 has
been end of life, only Sun 4170 is used in current system. But if Sun 4100 used in old
system is broken, you have to replace it with Sun 4170. During VRE Sun 4100
configuration, no OS disk is needed. But one specific OS disk for Sun 4170 is required
to be insert to the DVD driver in the host during VRE Sun4170 configuration.
VRF board contains an optical interface to receive scan data from Remote Receiver (RRx)
board, filters the data and DMAs the data to the ICN for image reconstruction. The VRF
board is physically installed in PCIs slot on the ICN. The VRF will receive an 80MHz
reference clock from the exciter board.
VRF is used in Sun 4100, and VRF2 is used in Sun 4170. The VRF2 is a redesign of the VRF
to change the FPGA from an Altera StratixGX device to an Altera ArriaGX device.
125
126
The SRFD3 RF Amplifier and Interface (SRFD3 Module) is the third phase in the development
of the Scaleable RF Driver (SRFD) subsystem. It is the first liquid cooled amplifier and lower
output power(10kW). The SRFD3 Module performs the following functions: Head/Body mode
switching, CAN communication link decoding, control and status communication between
the system and the RF amplifier, RF power amplification, and RF power sampling for power
monitoring by the Universal Power Monitor (UPM). The SRFD3 Module provides appropriate
test points, displays and adjustments to allow for calibration and troubleshooting in the field.
The SRFD3 Module will also incorporate water cooling for fan noise reduction.
The SRFD3 Module consists of five functional blocks: I/F Translation, Input Conditioning, RF
Amplifier, Output Conditioning and Input Power.
127
128
SRFD3 has been pre-adjusted in factory. You don’t need to adjust RF power on site during
installation.
Because the card 72 in the RF Power Measurement Kit is for 16KW power output, it is
necessary to modify the measurement method.
129
130
Performance:
Water cooling gradient system
Slew Rate: 100 T/m/s
G-Max: 33 mT/m
Input : 208VAC, 1A ; 200VDC&700VDC
Output: 300A for 100ms, 1400VDc
Cal/Function Check:
DC offset, DQA cal
Diagnostic:
Hammer, Frame/Clk test, ECC Verification, Hysteresis Static Test is obsolete.
FRU: XFA, XFD-PS, Control Boards
Install/Replacement:
1) Quick disconnecting design to facilitate the replacement.
2) Need to remove the leak sensor first to replace control board.
3) Super capacitor design of XFD-PS require ~30 minutes to discharge.
4) 87 Kg weight of XFD-PS. Need hoist tool
131
The XFD-PS is an IGBT-based switching Power Supply. It can output three sets of DC +200V,
+700V and -700V. The XFA needs the three kinds of voltages.
There are two serially connected 4.3 Farad supper capacitor (EDLC: Electrical dual layer
capacitor) in 200V DC power supply. By using super capacitor, the power of PDU can be
reduced.
132
The part is the 200V DC power supply for three XFAs and 200V to 700V DC-DC power supply.
133
134
Three sets of DC700V and -700V are isolated.
135
XFA is an IGBT-based switching amplifier. It provides a voltage to overcome the inductance
and resistance of the MR gradient coil. The XFD-PS powers the three XFAs.
Actually, the ideal output of XFA is a square wave. In order to decrease the time of rising
edge, XFA will output DC 1400V energized on the G-coil. After getting the top of square wave,
only 200V can sustain. The control board output the gate control signal to switch on or off
the IGBT to implement this function.
136
The XFD-PS weigh 87kg. Be strictly follow the replacement procedure in the service manual.
In case the connections are difficult to remove, follow Method B.
137
RF amplifier, gradient amplifiers and gradient power supply are all water-cooled in SV
system cabinet. Leak sensors are used to detect if there is leakage happened on any water
pipe connectors.
Leak sensor is composed of two parallel wires that terminated by a 8.3MOhms resistor at
one end. The electroconductive part of the wires are exposed every one centimeter. The
other end of the leak sensor are connected to the cabinet monitor.
• If the leakage sensor are broken, the resistance between A and B will be greater than
50MOhms. This will be labeled as SF (Sensor Failure) on the cabinet monitor.
• If any leakage happens, because the coolant resistance are less, the resistance between
A and B will less than 3MOhms. This will be labeled as LK (Leakage) on the cabinet
monitor.
• If there is no SF and LK, the resistance between A and B should be equal to 8.3MOhms.
According to three different resistance values, cabinet monitor can identify three status.
138
Leak sensors are classified three levels. Leak sensor 1 is the highest level, which is only used
in HDe system and will trip the PDU main breaker. In SV system, only Leak sensor 2 and 3 are
used.
Leak sensor 2 consists of three parts: Front, Bottom and Water Tray. Front sensor locates in
front of RF amplifier and XFD; Bottom sensor locates on the bottom of the system cabinet;
Water Tray sensor locates on the left side of tray. The three parts of sensor 2 are serially
connected and terminated on the water tray sensor.
Leak sensor 3 locates on the bottom of the water tray.
A fault on Leak sensor 2 will trip the E-Off.
A fault on Leak sensor 3 will produce an error message.
139
Function:
1. Cabinet monitor will trip the E-Off contactor in the PDU when
a) Leak Sensor 2 detects water leakage or are open
b) Temperature of system cabinet is over 45 degree centigrade
c) 4KW LCS or 8KW LCS/MCS stop working
2. Cabinet monitor only sends the error message to host through CAN link when
a) Leak Sensor 3 detects water leakage or are open
b) Temperature of system cabinet is over 40 and less 45 degree centigrade
c) Water level is low in any one of LCS or MCS
3. Cabinet monitor will trip the Night Mode contactor in the PDU when
The system haven’t scanned for one hour.
The Nigh Mode control signal is from IRF I/O board J23 to SCIF J101, then to SCIF J105, then to J10
in cabinet monitor.
4. Cabinet monitor will energize the Night Mode contactor in the PDU when
The system is running TPS reset or begins to scan.
5. Provide Isolated +24V DC power to whole CAN link.
Pump flow sensor and Tank level sensor are all Normally Closed switch, which means:
1. When LCS or MCS stop working, the pump flow sensor are close. And when they works normally,
the sensors are open.
2. When Tank level is low, the tank level sensor are close. And when the level is normal, the sensors
are open.
You can use DVM to check the resistance between the two sensor cables.
140
141
142
• PS for Mega SW & RRX provides three kinds of input voltage to Regulator Box through the
filters in the System Cabinet Interface Board(SCIF). But the cables from the SCIF
connectors to Regulator Box are very long, which will surely cause voltage drop on the
cables. The input voltage of Regulator Box must be within the required range with the
voltage drop. So the system uses two exactly identical cables and connectors to decrease
the voltage drop. The output power cables from Regulator Box to Mega SW are also two
exactly identical cables.
• Current SV system doesn’t have the individual 12V Power Supply, which
,is for future use
to support MNS. That is why 3 out of 4 input LEDs and 5 out of 6 output LEDs are lit in the
Regulator Box.
• There are five voltage regulators to transform three kinds of input voltage to five kinds of
output voltage. Voltage Regulator Board monitors all regulated voltages using
comparator chips and sends status information (Power Good) to Mega Switch. Only when
all five regulator outputs are normal, then Power Good signal is Ok, which is a DC signal
ranged in 2.2 to 3.3V at pin 19 of J2.
• In the Mega Switch, the regulator just transforms the 6.85V input voltage to other three
kinds of output voltage. Mega SW also monitors the output voltage and send status
information, and this status information and Power Good signal will be sent to a logic
AND gate to get the Mega SW Power Good signal. Only when output of Voltage Regulator
Box and Regulator in Mega SW are both normal, the Mega SW Power Good signal is OK.
But this signal is not sent to RRX in SV system.
• Other input voltages to Mega SW except 6.85V just go through the Mega SW and are
directly sent to RRX.
143
All the input, output and power good LED indicators can be observed through the heat
dissipation holes of the Regulator Box. Due to no power supply for MNS, Input LED DS12
and Output LED DS1 are off all the time, and other LED indicators are continuous lit .
144
Voltage Regulator Board monitors all regulated voltages using comparator chips and sends
status information to Mega Switch.
The output from the monitoring circuit is buffered and sent to Mega Switch as a 3.3V digital
signal of wired-AND logic.
Diagnostic information is provided on the front panel via LED’s.
145
146
The Mega Switch can be divided into ten modules:
1. RF Router: It is a multi-way switch, which routes the four RF inputs to one RF outputs.
2. Loopback Interface: It is used to distribute loopback signal from Exciter to all receive
chain for R1 calibration and diagnostics.
3. Mixer Module: It amplifies the input RF signal according to R1 gain and transform its
frequency to 16MHz.
4. Driver Module Interface: It is the interface to supply the DC bias, DC power to coils and
other control signals.
5. RRX Interface: It communicates with the RRX by outputting the 16MHz RF signals and
receiving control signal from RRX.
6. CPLD: It provides the interface of the control signals.
7. SRI3 Interface: It communicates with SRI by transferring Coil ID and other control and
status signals.
8. Clock Interface: It filters the clock signal and makes some transformation,, then sends
the clock signal to RRX.
9. TNS (Transient Noise Suppressor): It receives noise from outside antenna and compares
with threshold and output control signal to RRX.
10. Power Interface: It regulates one DC Voltage from Regulator Box for internal using. A
power supervisor monitors the regulated power. Other DC Voltages are transferred to
RRX directly.
147
1. The Mega Switch provides the RF routing circuit which routes the four RF inputs to one RF
outputs. The main functions of the RF router block are as following.
• Provide the receive Interface of RF signal
• Route the RF receive signals to proper down conversion channels.
• Provide protection circuitry at each RF signal input to protect the components from being
damaged by the transmit RF signal during scan, ESD(Electronic Static Discharge) in
installation and other occasions.
2. Loopback interface is used to distribute loopback signal from Exciter to all receive chain
for R1 calibration and diagnostics.
• R1 calibration is very important in SV system. It is performed when resetting TPS and
rebooting the system every time. Loopback signal is split into eight separate but equal
signals and are sent to each receive channel via SPDT(Single-Pole Double-Throw)
switches in the RF router, then to Mixer Module for amplifying and mixing, then output to
RRX.
• Meanwhile, system will set the R1 value from 13 to 1, and send them to RRX via DVMR
link. Then RRX sends the R1 data to Mixer Module in Mega Switch. Eight Mixer Modules
amplify or attenuate the loopback signal according to different R1 value.
• In VRE, system will check the signal power of every channel when R1 value is from 1 to
13, if the signal level is in the specification, R1 calibration is successful and system can go
to next step. But if it fails, system will report R1 calibration failure and stop.
• If RF disable switch is located in disable mode in Exciter, system will not get loopback
signal during TPS resetting rebooting, this will cause system fail.
148
149
4. Driver Module Interface has the following functions:
a) Multi Coil TR Bias
The Mega Switch receives Multicoil TR bias (MCB) from the Driver Module Lite and these signals are
routed to each coil port.
The MC TR bias is specified as following:
MC TR bias input (Transmit Mode) supplied by Driver Module MC Driver board +3, +5 or +7V +/- 0.1V @
500mA
Maximum MC TR bias input (Receive Mode) supplied by Driver Module MC Driver board -5 V +/- 0.1V @
500mA
b) Control Signal of RF Routing
The Mega Switch receives a 16 bit serial stream of data to control RF routing from the Driver Control
Board within the Driver Module on differential lines (DATA, STROBE, OE), converts these lines to digital
signals (data, strobe, enbale), and converts this serial data to parallel signals.
c) Power Supply From Driver Module
The Mega Switch receives +15V and –15V powers from the Driver Module Lite. These supply voltages
are used for each coil port, Body Preamp Bias Power. +10V power which is used for each coil port is
generated by linear regulators from +15V. Note that +10V is not supplied to the Legacy port, only +/-
15V are supplied.
d) Body Preamp Bias and Over Current Detect
The Body preamp inputs are biased with 15Vdc supply from the Driver Module through a inductor at
all times (Bias-T). This bias circuitry has PTC resettable fuse and detects over-current state, and
allows bias output current to 300 mA flows to the preamps at maximum. The intent of this method is
to protect the circuit traces and coaxial cables in the event of a shorted preamp or cable. A voltage
comparator circuit is provided to determine when such a fault condition has taken place. The
comparator’s output will remain Low so long as the bias circuit in over-current detected state. While
the comparator output is Low, yellow LED will remain illuminated.
150
6. The control signals from the Driver Module is decoded and the following signals are
generated.
• RF Router Control Signals (To RF Router Block)
• RRx Reset Signal (To RRx)
• External MUX Control Signals (To Coil Ports) : The CPLD just pass though this signal to each
Coil Port.
• Body Select Signal (To RF Router Block)
• Loopback Select Signal (To RF Router Block)
The control signals from the RRx Module is decoded and the following signal is generated.
• R1 Control Signals (To each Mixer Module)
• RCVR_UNBLANK
The CPLD provides the control signals to switch the receive path to loopback circuitry
when the UNBLANK signal is Logic Low (Blank state).
7. SRI3 interface consists of the following signals.
• COIL ID and HART
• COIL Present
• Body Transmit ENBL
• COIL LED
151
8. Clock Interface
The Mega Switch receives 80MHz Master Clock from Exciter. The clock signal is band pass filtered to
reduce the noise from the system, and routed to a LVPECL buffer to convert sine wave to LVPECL
signal. The converted clock signal is transmitted to the RRx module through TNS interface connector.
9. TNS (Transient Noise Suppressor)
The Mega Switch detects the possible occurrence of a significant transient noise event using a BNC
connected independent loop antenna. The RF signal from the antenna is bandwidth limited and
applied to a logarithmic (dB power to V) amplifier device, the output of which is compared to a
threshold provided by the RRx Module. 8bit D/A converter
,is used to generate the threshold level
voltage in the RRx module and the threshold level can be set to a value in the D/A Converter’s output
range. If the threshold is exceeded, a transient detect flag will be output to RRX.
10. Power Interface
a) Power Connector
The Mega Switch receives the supply voltage from the external power supply. The Mega Switch
provides a stacked SubD 37 pin female connector to receive 6V85_A, 6V85_D_RRx, N6V85_A,
5V05_A_RRx, 5V05V_D_RRx, 3V16_D_RRx and Power Good (PG) Signals. The pin assignment of the
power connector is the same to each other to for failsafe. The Mega Switch doesn’t use all of these
supply voltages mentioned above, just uses 6V85_A and N6V85_A. 6V85_A is regulated to 6V_A and
5V_A, and 5V_A is regulated to 3V3_D on the board, respectively. The received supply voltages are
distributed to the RRx Module and / or the MNS upconverter.
b) Power Supply Supervise
The Mega Switch provides the supervise circuit for the 6V regulator. The supervise circuit provides
the power good (PG) signal. Also, the Mega Switch has an interface to receive the PG signal from the
Regulator Box, which is the power supply board for the Mega Switch and RRx. The power good
signals are High in normal operation, but become Low when the regulated voltages become out of
the specified range. These two PG signals are fed into CPLD on the Mega Switch, and the CPLD
generates a signal which is AND logic output of these two PG signals. This signal from the CPLD is
passed to the RRx Module as the “ MegaSW_PG” signal.
152
153
The Remote Receiver’s main function is to provide analog and digital signal processing and
A/D conversion for 8 channels of data.
Key functions of the RRx Module are:
- 16-bit Analog-to-Digital conversion
- Digital decimation and filtering
- Synchronous transfer of 8 channels of 18-bit two’s complement I-data and Q-data from
the FPGA over the DVMR Common Communications link at sample rates of 500kHz (+/-
250kHz)
- Communication link to and from Megaswitch for diagnostic and control signals
- Control of TNS section of Megaswitch module and TNS notching for receiver channels
- Read of board type – needs to indicate an 8ch vs. 16ch RRx module
- Read of board status
154
LED DS2, DS3, DS4 used to identify the Receiver Module number in binary format. LED DS1 is
used to indicate when the Mega Switch +3.3V incoming power is good and is not lit when
either the Mega Switch power is down or the cable into connector J1 is not connected. DS1
is used to prevent a false RRX1 indicator, where Port ID is equal to ‘000’. All of these LEDs are
powered by the Mega Switch +3.3V power.
155
Pin function in Port A Connector:
A1, A2: Coil present
A3 ~ A10: M0~M4 signal from Mega SW to control multicoil
C1 ~ C4, D1 ~ D4: Ch1 ~ Ch8 preamp out, no bias, DC blocked
E1: Transmit signal up to 2KW for 1.5T or 4KW for 3.0T. Peak Power
E6: Reflected power to 50ohm load, 2.5KW Peak Power, 50W RMS
F1 ~ F8, G1 ~ G8: Multicoil T/R bias
F9, F10: To detect if this coil needs body coil to transmit
G9, G10: To detect if this coil needs system preamp to amplify
I2, I5: 15VDC for aux & preamp power
I3, I6: 10VDC for aux & preamp power
I4: -15VDC for aux power
I7, I8: Coil ID
I9, I10: Coil present’
156
• Fixed table has no Dock and Undock function like detachable table. Up and down of fixed
table are implemented by actuator, which is driven by electrical motor. The table can go
up and down very smoothly with very low noise compared to the detachable table. And
there is no other mechanism to drive fixed table up and down.
• The control box for fixed table is newly designed.
• The longitudinal movement of the cradle is driven by longitudinal motor located in the
rear pedestal. It is same as HDe.
• The cover of the cradle is reinforced to protect the embedded Express PA coil from
damage. The cradle with the Express PA coil moves the patient into and out of the
magnet bore.
157
There are five limit switches in the Fixed table.
1. One Home Sensor Limit Switch. Only when cradle moves to the home position, the Home
sensor Limit Switch will be activated to allow the fixed table to be moved up and down. It
can only be accessed for service by removing the top right cover.
2. Two Up Limit Switches and Two Down Limit Switches. (Shows the logic position diagram of
the four limit switch) This diagram shows the logical position relationship of the four limit
switches. Only Up1 Limit Switch locates on the fixed table frame. The other three limit
switches integrate in the actuator. When the table is going down, Down1 LS will be
touched first, then fixed table will stop if Down1 LS is OK. But if the Down1 LS doesn’t
work, the table will keep going down, Down2 LS will be touched and the table will stop. It
is same for Up1 LS and Up2 LS. In normal situation, only Up1 and Down1 will be touched.
Up2 and Down2 LS are redundant to assure the table can stop at UP and DOWN limit
position in case Up1 or Down1 LS are broken. The highest position of the fixed table can
be determined by adjusting the position of the UP LS Actuator. Only when the fixed table
is on the top height, the cradle can be moved in and out.
158
The table control board is a newly designed part, which is fixed on the dock frame shipped
with the fixed table package.
• PS for PHPS-LITE2 supplies DC 42V power as the input of table control board through SCIF
board. Then table control board outputs two direction power to the electrical actuator in
the fixed table. When the up or down foot pedal is pressed down, the table will move up
or down. And when the Up limit Switch or Down limit Switch is activated, the table will
stop.
• Table control box transfers the table vertical position information to SRI. When the table
is on the highest position, SRI will output Enable signal to PHPS Lite2 to enable the
longitudinal movement of the cradle. And it receives 8V power from the SRI for its logic
circuit working. The signal ground is connected to table control box shell through fasten
screw.
159
Please refer to the service method 5364191-2EN 'Replacement / Fix Table / TABLE
ELECTRICAL ACTUATOR' for the detailed procedure.
160
Express coil consists of three coils: HNA, AA and PA.
1. HNA and AA can’t be used together because of one port A.
2. HNA and PA can use together.
3. AA and PA can use together.
161
HNA coil consists of three parts: Anterior part, Posterior part and Horseshoe adapter.
There are 14 elements in HNA. Not all the elements will be used during scanning. Different
elements will be used in different scan mode.
HNA itself has four scan modes, which are Brain, NV, C Spine, C spine with adapter. And CT
Spine mode can only be available when combining HNA and PA.
162
AA coil consists of four elements. It can’t be used independently, and must be used with PA
coil, which their combination is functionally similar to 8ch Body Array or 4ch Torso Array.
They have three scan modes – BodyArray12, BodyArray_23 and BodyArray_34, which will
be shown on the next slide.
163
PA coil consists of 12 elements. In the surface of cradle, there are coil marker from 1 to 4.
Every one marker contains of 3 elements. They have three scan modes – SpineArray_12,
SpineArray_23 and SpineArray_34.
164
This table shows all the scan mode which the HNA, AA and PA have, and which elements will
be used in different mode.
165
There are three modes for Express coil MCQA. Phantom position and landmark position
would be strictly followed otherwise failure of MCQA. Be sure to use the foam pad in the
HNA when scanning HNA+PA mode and HNA with Adapter + PA mode. Please refer to
specific procedures of Express coil MCQA in the service manual.
166
• Automatic coil selection function can simplify the scan operation. The operator doesn’t
have to know exactly what scan mode should be selected. The system can automatically
select